Troubleshooting guide – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 24

Advertising
background image

20

13. TROUBLESHOOTING GUIDE

PROBLEM

CAUSE

SOLUTION

No display

Main power not switched on.

AC Adapter socket not inserted

properly.

Switch on the power supply.
Re-insert AC Adapter.

“Ur” (Under

range)
“Or” (Over

range)
in primary or

secondary

display

Measured value is out

of range.
Electrode not connected.
Electrode clogged, dirty

or broken.
Meter not calibrated.
Wrong temp value.

Check electrode is connected.
Clean or replace electrode.
Recalibrate the meter.
Confirm measurement condition.
Treat samples to bring within meter

measuring range.
Reset meter.

Slow response

Dirty electrode.
Temperature of sample is

changing.

Clean the electrode.
Allow temperature to stabilize.

Meter not

responding to

key press

Manual HOLD or Auto HOLD

is active.
Worn keypad.

Press

to deactivate manual

HOLD. See Section 9.4 to disable Auto

Hold feature if enabled.
Contact Technical Service.

Invalid key; Button is not

functional in the current

operation mode.

Press alternate key.

Select valid key depending on current

mode.

Secondary

display

continually

scrolls

The conductivity calibration

standard is not within 40% of

expected value during

automatic calibration.

Use fresh standard or use an alternate

calibration standard.

Ensure that the conductivity

electrode guard is attached.

The pH buffer is >1 pH unit

from the available pH buffers

remaining for calibration.

Use fresh standard or use an alternate

calibration standard. Verify the USA or

NIST buffer selection. Check pH

electrode.