2 single or multi-point calibration, Single or multi-point calibration – Thermo Fisher Scientific Eutech PC 700 User Manual

Page 11

Advertising
background image

7

If you only use calibration standards that are listed in

TABLE 1,

automatic calibration is recommended. If you intend to calibrate with

one or more standards that are not listed in

TABLE 1, the PC 700 must

be set for manual calibration instead.
The factory default is automatic conductivity calibration. See

Section

9.11

to change this setting. There is no automatic calibration value

available for conductivity range 1.

TABLE 1

Range

#

Conductivity

Range

Automatic Calibration Values

Normalization Temperature

25 ºC

20 ºC

r 1

0.00 – 20.00 μS

None

None

r 2

20.1 – 200.0 μS

84 μS

76 μS

r 3

201 – 2000 μS

1413 μS

1278 μS

r 4

2.01 – 20.00 mS

12.88 mS

11.67 mS

r 5

20.1 – 200.0 mS

111.8 mS

102.1 mS

Range

#

TDS Range

(using 0.5 TDS factor)

Automatic Calibration

Values

r 1

0 – 10.00 ppm

none

r 2

10.1 – 100.0 ppm

none

r 3

101 – 1000 ppm

none

r 4

1.01 – 10.00 ppt

none

r 5

10.1 – 100 ppt

none


5.2 Single or Multi-Point Calibration
Use Single-Point Calibration to apply a single calibration value across

all ranges.