2 single or multi-point calibration, Single or multi-point calibration – Thermo Fisher Scientific Eutech PC 700 User Manual
Page 11

7
If you only use calibration standards that are listed in
TABLE 1,
automatic calibration is recommended. If you intend to calibrate with
one or more standards that are not listed in
TABLE 1, the PC 700 must
be set for manual calibration instead.
The factory default is automatic conductivity calibration. See
to change this setting. There is no automatic calibration value
available for conductivity range 1.
TABLE 1
Range
#
Conductivity
Range
Automatic Calibration Values
Normalization Temperature
25 ºC
20 ºC
r 1
0.00 – 20.00 μS
None
None
r 2
20.1 – 200.0 μS
84 μS
76 μS
r 3
201 – 2000 μS
1413 μS
1278 μS
r 4
2.01 – 20.00 mS
12.88 mS
11.67 mS
r 5
20.1 – 200.0 mS
111.8 mS
102.1 mS
Range
#
TDS Range
(using 0.5 TDS factor)
Automatic Calibration
Values
r 1
0 – 10.00 ppm
none
r 2
10.1 – 100.0 ppm
none
r 3
101 – 1000 ppm
none
r 4
1.01 – 10.00 ppt
none
r 5
10.1 – 100 ppt
none
5.2 Single or Multi-Point Calibration
Use Single-Point Calibration to apply a single calibration value across
all ranges.