2 connection diagram – Thermo Fisher Scientific Alpha COND 2000 (In the process of updating the brand guideline & product image) User Manual

Page 17

Advertising
background image

Instruction Manual

αlpha CON 2000

11

3.2

Connection Diagram

Caution: Ensure electrical mains are disconnected before proceeding.

Connections for wall mounting version

L

N

PT100/ PT1000

Current

HOLD

OP2

Current

OP1

1

2

3

13

14

4

5

6

7

8

9

10

11

12

15

16

17

18

19

20

21

22

23

24

25

30

29

28

27

26

31

32

33

Anti-surge

PT1000

PT100/

1

L

2

N

Current

15

Current

13

OP2

14

16

OP1

17

3

4

5

6

22

HOLD

GND

18

19

21

20

23

24

25

7

8

9

10

11

29

26

28

27

30

31

32

33

12

250V, 315 mA,

Anti-surge

1.

AC mains live wire

18.

12V Power supply

2.

AC mains neutral wire

19.

Hold function

3.

AC mains protective earth wire

20.

Hold function

4.

Relay A (SP 1)

21.

Earth ground

5.

Relay A (SP 1)

22.

no connection

6.

Relay B (SP 2)

23.

no connection

7.

Relay B (SP 2)

24.

no connection

8.

Wash relay

25.

no connection

9. Wash

relay

26. Temperature

ground

10.

Alarm relay (NC)

27. Temperature

Sense

11.

Alarm relay common

28.

Temperature Input (short to terminal

12.

Alarm relay (NO)

27 if using 2- wire RTD)

13.

4-20 mA temperature output, +ve terminal 29.

no connection

14.

4-20 mA temperature output, -ve terminal

30. 4 Cell type Conductivity Input

15.

4-20 mA Conductivity output, +ve terminal **31. 4 Cell / 2 Cell type Conductivity Input

16.

4-20 mA Conductivity output, -ve terminal

**32. 4 Cell / 2 Cell type Conductivity Input

17. 12V

ground

33. 4 Cell type Conductivity Input

IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches

OFF under Alarm condition. Therefore the Alarm display device should be connected
to the ‘NC’ contacts of the relay (10 & 11).