Instruction Manual

αlpha CON 2000

30

6.9

Measuring range selection (RANG) sub-function

In this sub-function you select the measuring range.

SETUP

SETUP

SETUP

SETUP

SETUP

SETUP

SETUP

SETUP

HOLD

HOLD

HOLD

HOLD

HOLD

HOLD

HOLD

HOLD

µS

µS

µS

µS

mS

mS

mS

4

1

2

3

4

5

6

7

1

2

2

2

2

2

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

2

2

Measuring ranges:

No.

Measuring range

Resolution

1

0.000 to 2.000 µS/cm

0.001 µS/cm

2

0.00 to 20.00 µS/cm

0.01 µS/cm

3

0.0 to 200.0 µS/cm

0.1 µS/cm

4

0 to 2000 µS/cm

1 µS/cm

5

0.00 to 20.00 mS/cm

0.01 mS/cm

6

0.0 to 200.0 mS/cm

0.1 mS/cm

7

0 to 1000 mS/cm

1 mS/cm

Relative accuracy:

±1 % of full scale reading (±2 % >500 mS/cm)

System accuracy:

The effective measuring range and the total system accuracy depend on the

sensor used with the transmitter. Please refer to the respective sensor
documentation.


Press the ENT key to confirm your selection.
Continue with rest of the Setup procedures by pressing ▲ or ▼ key or return to the
Measurement mode by pressing the ▲ and ▼ key (escape) simultaneously.

1. Select the “RANG” sub-function and then press

the ENT key.

2. Selecting the measuring range: The upper

display shows the currently selected measuring
range and the lower display shows the cell

constant. The range number is displayed at
lower left corner.
Press the ▲ or ▼ key to select the requested
measuring range.