Current Output 2 (CUR 2) sub-function


In this sub-function you set the current output range of the transmitter for temperature values.

SETUP

HOLD

4

SETUP

HOLD

4

SETUP

HOLD

4

SETUP

SETUP

HOLD

HOLD

°C

°C

°C

°C

4

4

SETUP

SETUP

HOLD

HOLD

4

4

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1

2

2

3

3

4

4


1. Select the “CUR 2” sub-function and then press the ENT key.
2. Selecting the output type: press the ▲ or ▼ key to select the desired output type: 0-20 mA or

4-20 mA. Press the ENT key to confirm your selection.

3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA): Press

the ▲ or ▼ key to set the temperature value (setting range: -10.0 to 115.0 °C or 14 to 237 °F) to
be equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.

4. Setting the temperature value at which the transmitter output will be 20 mA: Press the ▲ or

▼ key to set the temperature value (setting range: 0.0 to 125.0 °C or 32 to 257 °F) to be
equivalent to 20 mA. Press the ENT key to confirm your setting.


Continue with rest of the Setup procedures by pressing ▲ or ▼ key or return to the
Measurement mode by pressing the ▲ and ▼ key (escape) simultaneously.