3 temperature setting (°c°f) sub-function – Thermo Fisher Scientific Alpha COND 2000 (In the process of updating the brand guideline & product image) User Manual

Page 27

Advertising
background image

Instruction Manual

αlpha CON 2000

21

6.3

Temperature setting (°C°F) sub-function

SETUP

SETUP

SETUP

HOLD

HOLD

HOLD

4

4

4

ENT

SETUP

SETUP

HOLD

HOLD

4

4

°C

SETUP

SETUP

SETUP

HOLD

HOLD

HOLD

4

4

4

ATC

°C

°F

ENT

ENT

ENT

ENT

ENT

ENT

1

2

2

3

3

4

5

6

ENT

1. Select the “SET °C°F” sub-function and then press the ENT key.
2. Selecting temperature unit: press the ▲ or ▼ key to select the desired temperature unit “°C” or

“°F”. Press the ENT key to confirm your selection.

3. Enable/disable ATC mode: press the ▲ or ▼ key to enable (ATC ON) or disable (ATC OFF)

automatic temperature compensation. Press the ENT key to confirm your selection.

ATC disabled (ATC OFF):

For manual temperature compensation you can set the process and calibration
temperatures. This allows calibration at a temperature other than the process

temperature. Example: setting a calibration temperature of 25°C lets you calibrate
using standard solutions at 25°C, even if your process temperature is different from

25°C.

4. Setting the process temperature: press the ▲ or ▼ key to adjust the process

temperature shown in upper display value (setting range: –10.0 to 125.0°C / 14.0
to 257.0 °F). Press the ENT key to confirm process temperature.

5. Setting the calibration temperature: press the ▲ or ▼ key to adjust the

calibration temperature shown in upper display value (setting range: – 10.0 to
125.0°C / 14.0 to 257.0 °F). Press the ENT key to confirm the calibration

temperature.

ATC enabled (ATC ON):

6. Setting the temperature offset: The display shows the current temperature

offset (upper display value) and the temperature currently measured (lower
display value). Compare the current temperature displayed on the transmitter to a

thermometer known to be accurate. Note the correct temperature value. Press

the ▲ or ▼ key to adjust the lower display value. The upper number indicates
the offset value. You can offset the temperature by up to ±10°C/±18°F. Press

the ENT key to confirm your setting.

Continue with rest of the Setup procedures by pressing ▲ or ▼ key or return to the

Measurement mode by pressing the ▲ and ▼ key (escape) simultaneously.