αlpha CON 2000

47



11.4.5 Control signal of Pulse Frequency Controllers

Relay

ON

tON

tOFF

Pulse Length T

OFF

Time [s]


The output relay of the pulse frequency controller is clock-timed. The pulse duration tON

remains constant at 250 msec. Depending on the divergence from the limit value, the
frequency (1/T) is increased or decreased in accordance with the proportional range Xp.
The following applies:
tON = Const. (250 msec.)
greater divergence Î greater f (greater frequency)
Xp exceeded Î max. frequency