Connections for panel mounting version

1.

AC mains live wire

**17. 4 Cell / 2 Cell type Conductivity Input

2.

AC mains neutral wire

18.

4 Cell type Conductivity Input

3.

AC mains protective earth wire

19.

12V Power supply

4.

Relay A (SP 1)

20.

12V ground

5.

Relay A (SP 1)

21.

no connection

6.

Relay B (SP 2)

22.

Earth ground

7.

Relay B (SP 2)

23.

no connection

8.

Wash relay

24.

Temperature ground

9.

Wash relay

25.

Temperature input

10.

Alarm relay (NC)

26.

Temperature sense (short to

11.

Alarm relay common

terminal 25 if using 2- wire RTD)

12.

Alarm relay (NO)

27.

no connection

13. Hold

function

28. 4-20

mA

temperature output, -ve terminal

14. Hold

function

29. 4-20

mA

temperature output, +ve terminal

15.

4 Cell type Conductivity Input

30.

4-20 mA Conductivity output, -ve terminal

**16. 4 Cell / 2 Cell type Conductivity Input 31.

4-20 mA Conductivity output, +ve terminal

IMPORTANT: The Alarm relay functions as an “Active Low” device i.e. it switches
OFF under Alarm condition. Therefore the Alarm display device should be connected

to the ‘NC’ contacts of the relay (10 & 11).

** When using 2 Cell type Conductivity electrode, terminal 15 should be shorted to
terminal 16 and terminal 18 should be shorted to terminal 17.