Instruction Manual

αlpha CON 2000

16

5

CALIBRATION MODE

You can access the Calibration mode directly from the Measurement mode by pressing the

CAL key and entering the Calibration security code “11”. Calibration mode may also be
accessed via the Setup mode (see section 6.1).

5.1

Entering Calibration mode

MEAS

SETUP

HOLD

µS

4

4

°C

SETUP

HOLD

4

Calibration

See section 5.2

CAL

ENT

ENT

ENT

ENT

ENT

While in Measurement mode press the CAL key (direct access) or ENT key (access via Setup mode).

• If CAL key was pressed: The display prompts you to enter the calibration security

code. Press the ▲ or ▼ key to set the calibration security code to “11”, then press
the ENT key
to confirm the calibration security code.

• If ENT key was pressed: The display prompts you to enter the security code. Press

the ▲ or ▼ key to set the security code to “22”. Press the ENT key to confirm the
security code, then press the ▲ or ▼ key to select the calibration sub-function.

The display shows “CAL CON”. To start calibrations refer to section 5.2.

Note: to exit Calibration mode at any time press the ▲ and ▼ keys
simultaneously
(escape). The transmitter returns to the Measurement mode and
the old calibration values remain active. The calibration security code automatically

resets from “11” to “000” after the transmitter returns to Measurement mode.