Configuration (CNFG) sub-function
In this sub-function you configure the transmitter to suit your requirements.
1. Select the “CNFG” sub-function and then press the ENT key.
2. Enabling/disabling the display backlight: Press the ▲ or ▼ key to switch display backlight on
(BKLT ON) or off (BKLT OFF). Press the ENT key to confirm your selection.
3. Setting the backlight intensity (appears only If backlight is on): Press the ▲ or ▼ key to set
backlight intensity (setting range: minimum 1 to maximum 4). Press the ENT key to confirm your
setting.
4. Selecting the cell type: Press the ▲ or ▼ key to select “2-cell” or “4-cell” type. Press the ENT
key to confirm your selection.
5. Setting the alarm delay time in seconds: Press the ▲ or ▼ key to set the alarm delay time
(setting range: 0 to 2000 seconds). Press the ENT key to confirm your setting.
Note: With this parameter you set the delay before the alarm relay activates when
the set point has been exceeded.
SETUP
SETUP
HOLD
HOLD
4
4
SETUP
SETUP
SETUP
HOLD
HOLD
HOLD
4
4
4
SETUP
SETUP
HOLD
HOLD
4
4
SETUP
HOLD
4
SETUP
HOLD
4
4
4
4
SETUP
HOLD
4
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
ENT
MEAS
°C
ATC
1431
4
µS
1
2
2
3
4
5
4
6
6
7
8
8
8