Wash relay (WASH) sub-function


In this sub-function you set the parameters for the wash relay.

SETUP

SETUP

HOLD

HOLD

4

4

SETUP

SETUP

SETUP

HOLD

HOLD

HOLD

4

4

4

SETUP

SETUP

HOLD

HOLD

4

4

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1

2

2

3

3

4

5


1. Select the “WASH” sub-function and then press the ENT key.
2. Enabling/disabling wash function: Press the ▲ or ▼ key to enable (WASH ON) or disable

(WASH OFF) wash function. Press the ENT key to confirm your selection.

− If “WASH OFF” was selected, the transmitter returns “WASH” sub-function.
− If “WASH ON was selected:

3. Selecting the relay status condition: Press the ▲ or ▼ key to choose desired relay

status (de-energised = “DEEN” or energised = “EN”). Press the ENT key to confirm your

selection.

4. Setting the wash interval in hours: Press the ▲ or ▼ key to set the desired wash

interval (setting range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.

5. Setting the wash duration in seconds: Press the ▲ or ▼ key to set the desired wash

duration (setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.

Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the
HOLD function is activated 5 seconds prior and deactivated 10 seconds after the

wash cycle.



Continue with rest of the Setup procedures by pressing ▲ or ▼ key or return to the
Measurement mode by pressing the ▲ and ▼ key (escape) simultaneously.