2assembly and installation – Thermo Fisher Scientific Alpha COND 1000 User Manual
Page 10
![background image](/manuals/370681/10/background.png)
Operating Instructions
Thermo Scientific Alpha COND 1000
2
2
ASSEMBLY AND INSTALLATION
2.1
Measurement and Control System
A typical measurement system consists of:
•
a Alpha COND 1000 process controller
• a suitable Conductivity electrode with the appropriate Cell constant and
integrated temperature sensor Pt 1000 or Pt 100,
•
an immersion, flow or process assembly
•
a final control element such as pump or valve and
•
a chart recorder.
Power Mains
(220/110 VAC)
Alpha-CON1000 Controller
Flow Assembly
to Dosing Pumps
Process Assembly
with Electrode
Chart Recorder
Pt100/Pt1000
Temperature Sensor
Alpha-CON1000
Alarm / Siren
System
2.2 Unit
Dimensions
The field-tested control panel housing is 96 x 96 mm; with protection class IP 54
(front).
max. 175
max. 45
Mounting Cut-Out
Flat Gasket (1mm)
(To be Inserted By Customer)
Note: The Taped Corners Have to Be On Top
92
32
56
96
92 + 0.5
92 + 0.5
Alpha COND 1000