Thermo Scientific Alpha COND 1000

2

2

ASSEMBLY AND INSTALLATION

2.1

Measurement and Control System

A typical measurement system consists of:

a Alpha COND 1000 process controller

• a suitable Conductivity electrode with the appropriate Cell constant and

integrated temperature sensor Pt 1000 or Pt 100,

an immersion, flow or process assembly

a final control element such as pump or valve and

a chart recorder.

Power Mains

(220/110 VAC)

Alpha-CON1000 Controller

Flow Assembly

to Dosing Pumps

Process Assembly

with Electrode

Chart Recorder

Pt100/Pt1000

Temperature Sensor

Alpha-CON1000

Alarm / Siren

System

2.2 Unit

Dimensions

The field-tested control panel housing is 96 x 96 mm; with protection class IP 54
(front).

max. 175

max. 45

Mounting Cut-Out

Flat Gasket (1mm)

(To be Inserted By Customer)

Note: The Taped Corners Have to Be On Top

92

32

56

96

92 + 0.5

92 + 0.5

Alpha COND 1000