6) After you set the process temperature value, press the ENTER key. The upper

display shows the current calibration temperature and the lower display shows
“C.ºC” to indicate calibration temperature.

7) Press

the

or keys to adjust the calibration temperature value. You can

adjust the value from -9.9 to 125 ºC.

8) Press

ENTER key to confirm your selection.

9) Continue with additional Additional Set-up procedures, or return to

Measurement mode by pressing the

and keys (escape) simultaneously).


7.3

Control Relay A/Control Relay B (SP1/SP2) sub-function

The SP1 option sets the operating parameters for Relay A; and SP2 for relay B.
Since these groups have the same set-up parameters, they are described together.

6

6

6

6

6

Sp1

Sp1

Hys

O .

OF.

μ

S

μ

S

HOLD

HOLD

HOLD

HOLD

HOLD

HOLD

SETUP

SETUP

SETUP

SETUP

SETUP

SETUP

Sp1

100

L

20

0

0



7.3.1

Entering the Set point 1 (or Set point 2) sub-function

The alpha CON 1000 controller/transmitter can operate in either one of the 10
different conductivity ranges (see Section 12.2 – Appendix 2: Measurement Ranges
Available in the Controller.
Refer to Section 7.5.1 for Selecting the Measuring
Range sub-
function.

1) Enter Advanced Set-up mode. Push the ENTER key and

(use

or keys to)

scroll to Advanced Set-up

security code “22”. Push the ENTER key again.

2) Press

the

or keys to scroll until the upper display

shows SP1 (or SP2).


7.3.2

Selecting the set point values

This lets you choose the value that will cause your controller
to activate (Default: SP1 = 100 µS; SP2 = 1900 µS).

1) Press

the ENTER key until upper display shows the

current set point value and the lower display shows SP1
(or SP2).

2) Press

the

or keys to select your value for Set point 1

(or Set point 2). Your controller will activate at the value
you select.

3) Press

the ENTER key to confirm your selection.

4) Proceed to 7.3.3, or return to Measurement mode by

pressing the

andkeys simultaneously (escape).

6

6

Sp1

Sp1

μ

S

HOLD

HOLD

HOLD

SETUP

SETUP

SETUP

Sp1

100

L