Operating Instructions

Thermo Scientific Alpha COND 1000

21

7.5

Measurement Range sub-function

In this sub-function, the appropriate range is
selected with the appropriate cell constant.

7.5.1 Selecting

the

Measuring Range sub-

function

1) Enter Advanced Set-up mode. Push the ENTER

key and scroll to Advanced Set-up security code
“22”. Push the ENTER key again.

2) Press

the

or keys to scroll until the upper

display shows “rng”.

3) Press

the ENTER key. The upper display shows

the maximum measurement range, while the lower
display shows the Cell constant. In the lower left-corner of the LCD, the number
corresponding to the respective range is displayed.

4) Press

the

orkeys to select the correct range. (Please refer to Section

12.2 Appendix 2 for the full list of measurement ranges).

5) Proceed to 7.6 to set current output or return to Measurement mode by

pressing the

andkeys simultaneously (escape).