TECHNICAL SPECIFICATIONS

Conductivity Range

Resolution

Default Cell
Constant, K

Optional Cell, K

0.000 to 1.999

μS/cm 0.001

μS/cm

0.01 -

0.00 to 19.99

μS/cm 0.01

μS/cm

0.01 0.1

0.0 to 199.9

μS/cm 0.1

μS/cm

0.1 1.0

0 to 1999

μS/cm 1

μS/cm

1.0 -

0 to 5000

μS/cm 5

μS/cm

1.0 -

0.00 to 19.99 mS/cm

0.01 mS/cm

1.0

-

0.0 to 199.9 mS/cm

0.1 mS/cm

10.0

1.0


Temperature 0-50

o

C (Display: -9.9 to 125

o

C)

Resolution

0.1

o

C

Relative Accuracy

± 0.5

o

C

Sensor

Pt 1000/Pt 100

Temperature Compensation

Auto / manual (reference at 25.0

o

C)

Set-point and Controller Functions
Controller characteristics

Limit / proportional (pulse or frequency) controller

Pickup / Dropout delay

0 to 1999 sec.

Switching Condutivity hysteresis

0 to 10% of Full Scale

Contact outputs, controller

2 potential-free change-over contacts

Switching voltage

max. 250 VAC

Switching current

max. 3A

Switching power

max. 600 VA

Alarm Functions

Function (switchable)

Latching / pulse

Pickup delay

0 to 1999 sec.

Switching voltage

max. 250 VAC

Switching current

max. 3A

Switching power

max. 600 VA

Electrical Data and Connections

Power Requirements

110 / 220 VAC (jumper selectable)

Frequency

48 to 62 Hz

Signal Output

0/4 to 20 mA, galvanically isolated

Signal Output Load

max. 600

Ω

Connection terminal

Terminal blocks 5-pole / 17-pole, removable

Mains fuse / fine wire fuse

slow-blow 250 V / 100 mA

EMC Specifications

Emissions

According to EN 50081-1

Susceptibility

According to EN 50082-1

Environmental Conditions
Ambient temp. operating range

0 to 50

o

C

Relative humidity

10 to 95%, non-condensing

Mechanical Specifications
Dimensions (control panel housing -
L x H x W)

175 x 96 x 96 mm

Weights (control panel housing)

max. 0.7 kg

Material

ABS with polycarbonate (front housing)

Insulation (Front / Housing)

IP 54 / IP 40