Operating Instructions

Thermo Scientific Alpha COND 1000

30

11.4

Guidelines for Returning Unit for Repair

Use the original packaging material, if possible when shipping the unit for repair.
Otherwise wrap it with bubble pack and use a corrugated box for better protection.
Include a brief description of any faults suspected for the convenience of Customer
Service Dept., if possible.

12 APPENDICES

ENSURE that the power cable is physically separated from the power
supply

.

12.1 Appendix 1 – Jumper Positions

Jumper Positions - Internal to the controller

JP 1

Selects the input voltage 220 VAC.

JP 2

Selects the input voltage 110 VAC.

JP 6

Selects the temperature sensor for Pt1000/Pt100

Fuse

Note that there is a fuse (slow-blow 100mA) internal to the
controller. Replace fuse with the recommended type only.

JP6 On the Analog PCB:
Selects between Pt100 and Pt1000. To switch between Pt100 and Pt1000, use
tweezers or needle nose pliers to shift jumper from Position A to Position B.

View from the top of Main PCB

F

ro

n

t

R

e

a

r

1

1

0

V

A

C

JP2

2

3

0

V

A

C

JP1

Fuse

View from the front of the Analog PCB

JP6

JP6

B

A

A - Pt100

B - Pt1000

Analog PCB