Operating Instructions

Thermo Scientific Alpha COND 1000

23

7.6.4

Selecting Conductivity value at 20mA

This parameter lets you choose the conductivity value at which the transmitter
output will be 20mA.

1) Follow directions in 7.6.1 to enter Current Output mode.

2) Press

the ENTER key. Scroll with the

orkeys until the upper display

shows a conductivity value and the lower display shows “r.20”.

3) Press

the

or keys to select the required conductivity value to be

equivalent to 20 mA (20.0 to 100.0% F.S.; Default is 100% F.S.).

4) Press

the ENTER key to confirm your selection.

5) Press the ENTER key to return to Advanced Set-up mode, or return to

Measurement mode by pressing the

andkeys simultaneously (escape).

7.7

Configuration (ConF) sub-function

This group of parameters lets you configure the controller to suit your requirements.

7.7.1

Entering the Configuration sub-function

1) Enter Advanced Set-up mode. Push the ENTER key and scroll to Advanced

Set-up security code “22”. Push the ENTER key again.

2) Press

the

or keys to scroll until the upper display shows “ConF”. Press

ENTER key to enter configuration sub-function.


7.7.2

Selecting Filter Function and the Alarm or Wash Function

In this configuration sub-function, you have the ‘Filter’ selection
which lets you filter rapid measurement changes and stabilising
the measurement. This configuration sub-function also allows
you to use the Alarm relay as Wash or Clean contact. The Wash
(Clean) contact is used in combination with automatic cleaning
systems. During the wash cycle, the analog output is set on hold.

1) In Filter selection mode, use the arrow keys to choose filter

"on" or "off".

2) Press the ENTER key to confirm. Display will now show

“ALr” or “CLn”.

3) Press

the

▲ or ▼key to choose "ALr" (Alarm) or "CLn"

(Clean) function.