7.3.5

Setting an on-delay time lag

You can set as time delay for each relay, which stops the relay from switching on
the moment the set point is exceeded. This controller lets you set a 0 to 1999
seconds time delay before the relay activates.

1) Follow directions in 7.3.1 to enter Control Relay

mode.

2) Press

the ENTER key. Scroll with the

and

keys until the upper display shows “0” time and the
lower display shows “On.d”.

3) Press

the

or keys to enter on-delay time for

Set point 1 (Set point 2). The controller will delay
activation for the number of seconds (0 to 1999) you select.

4) Press

the ENTER key to confirm your selection.

5) Proceed to 7.3.6, or return to Measurement mode by pressing the

and

keys simultaneously (escape).


7.3.6

Setting an off-delay time lag

You can set as time delay for each relay, which stops the relay from switching off
the moment the value reached the set point and hysteresis. This controller lets you
set a 0 to 1999 seconds time delay before your relay deactivates.

1) Follow directions in 7.3.1 to enter Control Relay

mode.

2) Press

the ENTER key. Scroll with the

orkeys

until the upper display shows “0” time and the lower
display shows “OF.d”.

3) Press

the

or keys to enter on-delay time for

Set point 1 (Set point 2). Your controller will delay
activation for the number of seconds (0 to 1999) you select.

4) Press

the ENTER key to confirm your selection.

5) Continue with Advanced Set-up mode procedures, or return to Measurement

mode by pressing the

andkeys simultaneously (escape).