5.2 Security

Codes

This controller has two levels of security protection with separate security codes.
The first level allows entry into the Calibration mode: security code = 11; the second
allows entry into the SETUP mode: security code = 22.

The security codes protect the controller from unauthorised tampering of its current
setting. The parameters cannot be changed unless the security code is entered.

5.2.1

How to enter and change parameters in Calibration mode

1) Press the CAL key. The upper display shows “000”

and the lower display shows “C.Cd” to prompt the
user to enter the Calibration security code.

2) Press the

or keys to scroll upper display to

Calibration security code “11”.

3) Press the ENTER key. The display shows “CAL

CON”.

4) Press ENTER key again to begin calibration. Refer to Section 6 for full details on

calibration.

5) Press the

or keys simultaneously (escape) to return to the Measurement

mode.

NOTE: To view (not change) the SETUP parameters, push the ENTER key
when the security code reads “000”.


5.2.1.1

Clearing the Calibration security code from the display

The calibration security code automatically resets from “11” to “000” after you return
to Measurement mode, so you do not need to clear the security code from the
display.

5.2.2

How to enter and change parameters in Advanced Setup mode

1) Press the ENTER key once. The upper display shows “000” and the lower

display shows “S.Cd” to prompt you to enter the Advanced Setup security code.

2) Press the

or keys to scroll the display to Setup security code “22”. NOTE:

Pressing the ENTER key at a value other than “22” causes the controller to
revert to the Measurement mode.

3) Press the ENTER key.
4) The upper display reads “tc”.
5) You are now in the Advanced Setup mode. See Section 7 for complete

instructions. To return to Measurement mode, press the

or keys

simultaneously (escape).
NOTE: If you want to view (not change) set up parameters, push the ENTER
key when the security code reads “000”.