Operating Instructions

Thermo Scientific Alpha COND 1000

20


7.4.4

Selecting Proportional Range Value, Xp

Note: If the controller type “oFF” is set, the parameters listed in 7.4.3, 7.4.4, and
7.4.5 are blanked out.

This mode lets you set a band as a percentage of its full scale value. You can select
this range from 10 to 200%, and the lower display shows “PrP”.

1) Follow directions in 7.4.1 to enter Controller mode.

2) Press the ENTER key. Follow directions in 7.4.2 and

select either “PLC” (pulse length control) or “PFC”
(pulse frequency control). Press ENTER key.

3) Press the ENTER key until the upper display shows the

proportional range (a number from 10 to 200%), and the
lower display shows “PrP”.

4) Press the

orkeys to choose the proportional range value Xp.

5) Press the ENTER key to confirm your selection.

6) Proceed to 7.4.5, or return to Measurement mode by pressing the

▲ and ▼

keys simultaneously (escape).


7.4.5

Maximum Pulse Length (tPL) or Maximum Frequency (FPF)

Note: If the controller type “oFF” is set, the parameters listed in 7.4.3, 7.4.4, and
7.4.5 are blanked out.

This mode lets you set the maximum pulse length or the
maximum frequency at which the relay will operate.

1) Follow directions in 7.4.1 to enter Controller mode.

2) Press the ENTER key. Scroll until the lower display

shows “t.PL” or “F.PF”.

a. In PLC (pulse length) mode: The lower display

shows “t.PL” to indicate pulse length. The upper
display shows your current pulse length. You
can select any value from 0.5 to 20 seconds.

b. In PFC (pulse frequency) mode: The lower

display shows “F.PF” to indicate pulse
frequency. The upper display shows your
current maximum pulse rate. You can select any value from 60 to 120
pulses per minute. When the measured value exceeds the Proportional
Band in 7.4.4, the controller will pulse the relay at this rate.

3) Press the

▲ or ▼ key to choose the period duration or maximum frequency,

depending on your mode.

4) Press the ENTER key to confirm your selection and to return to Advanced Set-

up mode, or return to Measurement mode by pressing the

▲ and ▼ keys

simultaneously (escape).

Note: The alarm contact is always of the fail-safe type. In the event of power failure
the alarm is triggered.