A25.0, Ap. c c. c – Thermo Fisher Scientific Alpha COND 1000 User Manual

Page 22

Advertising
background image

Operating Instructions

Thermo Scientific Alpha COND 1000

14

7.2

Temperature Calibration (ATC mode only)

1) Enter Advanced Set-up mode. Push the ENTER key and

(use

or keys to) scroll to Advanced Set-up security

code “22”. Push the ENTER key again.

2) Press

the

or keys to scroll until the display shows

“SEt ºC”. Push ENTER key.

3) Press

the

or keys to select “on ATC”. Press ENTER

key.

4) The upper display indicates the current temperature offset.

The current measured temperature is shown in the lower
display.

5) Compare the current measured temperature on the

controller display to a thermometer known to be accurate.
Note down the correct temperature.

6) Press

the

or keys to scroll the lower display to

match the correct value. The upper display will now show
the offset value. You can offset temperature up to ± 5 ºC.

7) Press

the

ENTER key to confirm your selection.

8) Continue with additional Advanced Set-up procedures, or return to the

Measurement mode by pressing the

and keys (escape) simultaneously).


7.2.1

Setting Manual Temperature Compensation

NOTE: This parameter is blanked out when the controller is
set for ATC operation.

For manual temperature compensation, you can set two
different temperatures: Process and Calibration. This allows
calibration at a temperature other than your process
temperature. Example: setting a calibration temperature of 25
ºC lets you calibrate using standard buffer solutions at 25 ºC,
even if your process temperature is at different temperature.

1) Enter Advanced Set-up mode. Push the ENTER key and

(use

or keys to) scroll to Advanced Set-up security

code “22”. Push the ENTER key again.

2) Press

the

or keys to scroll until the display shows

“SEt ºC”. Push ENTER key.

3) Press

the

or keys to select “oFF ATC”. Press

ENTER key.

4) The upper display shows the current process temperature

and the lower display shows “P. ºC” to indicate process temperature.

5) Press

the

or keys to adjust the process temperature value. You can

adjust the value from -9.9 to 125 ºC.