Operating Instructions

Thermo Scientific Alpha COND 1000

25

7.7.5

Wash contact (if the relay 3 is set to Wash)

1) Follow directions in 7.7.1 to enter Configuration mode.

2) Press

the

ENTER key until display shows “ALr” or “CLn”.

Choose “CLn” by pressing

▲ or ▼ key. Press the

ENTER key.

3) Press

the

▲ or ▼ key to select the wash cycle (interval

0.1 to 199.9 hours) and press the ENTER key to confirm.

4) Press

the

▲ or ▼ key to select the wash duration (1 to

1999 seconds) and press the ENTER key to confirm.

5) Proceed to 7.7.6, or return to Measurement mode by

pressing the

▲ and ▼ keys simultaneously (escape).


7.7.6

Input Line Resistance Adjust

This function compensates for the line resistance of the cable to its cell.

1) Follow directions in 7.7.1 to enter Configuration mode.

2) Press the ENTER key. Scroll with the

or keys

until the upper display shows “0.0” and lower display
shows “L.Ad”.

-

L.Ad = line adjuster resistance (0.0 to 100.0)

3) Press

the

orkeys to input the value.

4) Press

the ENTER key to confirm your selection.

5) Proceed to 7.7.7, or return to Measurement mode by pressing the

and

keys simultaneously (escape).


7.7.7

Reverting to factory default settings

Use this parameter to reset all settings to factory default.
Changing from “no” to “YES” and pressing the ENTER
key resets all settings to factory default.

WARNING: If you select yes, all the settings you have
made will be overwritten as a result!

1) Follow directions in 7.7.1 to enter Configuration mode.

2) Press

the ENTER key. Scroll with the

orkeys

until the upper display shows “no” or “YES”, and the lower display shows “deF”
(default).

3) Press

the

or keys to select no or yes. Selecting “YES” and pressing the

ENTER key will overwrite all setting you have made and automatically return
you to Measurement mode.

4) Press

the ENTER key to confirm your selection and to return to Advanced Set-

up mode, or return to Measurement mode by pressing the

andkeys

simultaneously (escape).

HOLD

SETUP

0.0

L.a

HOLD

SETUP

EF

HOLD

HOLD

SETUP

SETUP

10.0

10

i