6 extended operating level 20, Kg дььь – West Control Solutions SG 45 User Manual

Page 20

Advertising
background image

5.4.4

Slave pointer function

The minimum and maximum input values are stored in the unit.

The minimum input value is displayed as long as key Ì is
pressed.

The maximum input value is displayed as long as key È
is pressed.

Deleting the minimum value
The minimum value is deleted by pressing key È whilst key Ì is kept pressed.
Whether the minimum value should be deleted also by the digital input (

rES.L) can be determined during configura-

tion.
Deleting the maximum value
To delete the maximum value, press key Ì whilst keeping key È pressed.
Whether the maximum value should be deleted also by the digital input (

rES.H) can be determined during configura-

tion.

Deleting the minimum and maximum values is possible also via interface.

g

When de-energizing UNIFLEX SG 45 the minimum and maximum values are deleted.

g

In case of error of the display value (e.g. input fail behaviour), the minimum and maximum values are also
set to FAIL. When a valid value is displayed again, the minimum and maximum value are deleted.

5.4.5

Selecting the units

The unit to be displayed is determined via configuration

D.Unt.

By selecting

D.Unt = 22, display of any max. 5-digit unit or text can be determined.

1

Unit (example): kilogram

2

Text (example): TAG no.

Operation

UNIFLEX SG 45

Displays in the operating level

20

1 2 F

E

450.3

502.4

дььь

1 2 F

E

450.3

ыC

дььь

max

1 2 F

E

450.3

26.7

дььь

1 2 F

E

450.3

ыC

дььь

min

1 2 F

E

4.5

Kg

дььь

1 2 F

E

450.3

TI451

дььь

1

2

Advertising