Badger Meter M-Series M3000 User Manual

Page 25

Advertising
background image

25

Set Flow Direction
1. At MEASUREMENTS 20 press + fi ve times to set the

arrow at Flow Dir. Press E.

2. At FLOW DIR 25 press + or - to select Uni-Directional or

Bi-Directional.

3. This type of screen shows for 2 seconds.

Set Damping Factor
1. At MEASUREMENTS 20 press + six times to move the

arrow to Damping Factor. Press E.

2. At DampingFactor 26 press Exit this Menu, No Damping

or press a Time Frame.

3. If No Damping is selected, this type of screen shows for

2 seconds.

4. And this type of screen shows for 2 seconds when

Damping Factor time is selected.

5. MEASUREMENTS 20 screen returns with the arrow
pointing at Exit this Menu.

6. Press E once to place cursor at Main Menu 00.

7. Press E once to move cursor back to the Main Screen.

Empty Pipe Calibration

1. Press E once to move to Main Menu.

2. In the MAIN MENU 00, select Meter Setup.

3. In METER SETUP 10, select Empty Pipe.

4. In EMPTY PIPE 15, select Cal. empty pipe. Next, press E.