Use the sling-rigged method to lift large detectors into a
vertical position while still crated. Use this method to position
large detectors vertically into pipelines.

Do not lift a detector with a forklift via the detector body between
the fl anges. The housing could be dented and/or damage caused
to internal coil assemblies.

NEVER place forklift forks or rigging chains, straps, slings,

hooks or other objects inside or through the detector fl ow

tube for lifting or handling purposes. The isolating liner could

be damaged, rendering the unit inoperable.

Instructi

ons Specifi c to Hazardous Area Installations

These instructions apply to equipment covered by FM
Certifi cate Number IAB-182-01.

1.

The temperature range for fl uids passing through the
detector is -4°F to +248°F (-20°C to +120°C).

2.

The ambient temperature range surrounding the
amplifi er is -4°F to +122°F (-20°C to +50°C).

3.

The ambient temperature range surrounding the junction
box must not exceed 122°F (50°C).

4.

During any installation or repair, perform all procedures in
accordance with the applicable code of practice.

5.

Suitably trained personnel shall perform all installation
or repair procedures.

6.

If the equipment is likely to come into contact with
aggressive substances, it is the responsibility of the
user to take suitable precautions that prevent it from
being adversely affected, thus ensuring that the type of
protection is not compromised.

Aggressive Substances – e.g. acidic liquids or gases

that may attack metals, or solvents that may affect
polymeric materials

Suitable Precautions – e.g. regular checks as part of
routine inspections or establishing, from the material’s
data sheet that it is resistant to specifi c chemicals

Additional Information
Certifi cation markings are noted on the product label.
Markings include:

Mounted amplifi er 85-240 VAC

Remote mount 85-240 VAC