M5000 Harness Connections

1. Clip the alligator clip labeled Detector Ground to either of the hex nuts mounted on top of the meter flanges.

2. Plug the connector labeled Amplifier Electrode into the circuit board connector labeled "E1,

, E2,

, EP.

3. Plug the Amplifier Coil Output into the circuit board connector labeled C1, C2.
4. Plug the Output 1 into the board output connector labeled Out1.
5. Plug the Output 2 into the board output connector labeled Out2
6. Plug the Output 3 into the board output connector labeled Out3.
7. Plug the Output 4 into the board output connector labeled Out4.
8. Plug the RS232 into the board output connector labeled RS232.
9. Connect the harness wire connector labeled Detector Electrode to the 5-wire connector from the detector.
10. Connect the harness wire connector labeled Detector Coil to the 2-wire connector from the detector.

Figure 8: M5000 Wire Harness Circuit Board Connectors

Detector Electrode

Detector Coil

Detector Ground

Figure 9: M5000 Detector Connections