Manual tests – Badger Meter M-Series Field Verification Device User Manual

Page 17

Advertising
background image

Page 17

8-11

User Guide

9. The testing is completed automatically in 10 steps. During the test the flow meter shows Testing in progress on the

display.

10. The result is Passed or Failed.

11. If the test Failed, press the function key View Report on the upper left to see the results. See the example below.

Manual Tests

The result of the manual tests are not stored in the memory of the Veriification Device and cannot be downloaded to the PC

program.

1. Switch off the flow meter and connect the specific wire harness to the amplifier circuit board
2. Connect the male D-25 connector of the harness to the corresponding female connector on the Verification Device.
3. Switch on the flowmeter and be sure that the meter is not in the programing mode when the test is started.
4. Press the On/Off button on the Device and wait for the SelfTest to complete.
5. When the Start Menu appears in the display, press the upper left function key.
6. Select the menu Flow Meter and press the OK button.