Time, Contrast, Flow meter modbus address – Badger Meter M-Series Field Verification Device User Manual

Page 15

Advertising
background image

Page 15

8-11

User Guide

Time

1. Select StartMenu > Settings > Misc > Time.
2. Edit the hour and minutes in the edit box by using the numeric keypad. Use the right arrow key to move the cursor.
3. Confirm the new date with the upper right function key..

Contrast

Adjust the contrast of the display via the function keys (arrows) and save the adjustment with Yes.

Flow Meter Modbus Address

1. Select StartMenu > Settings > FM Modbus Address.
2. Edit the Address in the edit box by using the numeric keypad. Use the Delete arrow key to remove the last number

position.

3. Press the upper right function key to confirm the new address.
4. Be sure that the flow meter is programmed with the same modbus address or communication will fail. The default

address is 1.