Field Verification Device

Field Verification Device Functions

The Device also:

•

Determines if the M-Series meter is within one percent of the original factory calibration.

•

Verifies the functionality of all of the meter's inputs and outputs.

•

Measures electrode resistance and integrity.

•

Measures coil resistance and integrity.

•

Measures coil insulation resistance.

•

Measures current and frequency at previously selected flow rate.

•

Evaluates the signal processing functionality.

•

Verifies that there is no high-voltage noise disturbing the coil circuit.

Verification Device Kit Components

The Field Verification Device is packaged in a foam-lined, durable plastic case and includes the following components.

1. One (1) Verification Device +5 VAC, 3.0A power adapter
2. Four (4) AC power conversion connectors
3. One (1) USB PC data cable
4. One (1) DC power adapter
5. Two (2) Verification cable harnesses: one for the M2000 and one for the M5000

Figure 1: Kit Components