E-m-pf4-v1_00 – ROTRONIC PF4 User Manual

Page 10

Advertising
background image

E-M-PF4-V1_00

Rotronic AG

Bassersdorf, Switzerland

Document Code

Unit

PF4 Differential Pressure Transmitter:

Instruction Manual

Instruction Manual

Document Type

Page 10 of 54

Document Title

© 2014; Rotronic AG

E-M-PF4-V1_00

3.2.3

Filters

The curve of the differential pressure measured values can be smoothened with the help of a low-pass filter

with selectable time constant. There are 13 smoothing filter stages available. Practice has shown stages 0...7

meet most application requirements.

Filter Stage

Time Constant

63

Time Constant

99

Description

0

0 s

0 s

No filtering

1

0.4 s

1.8 s

Weak filtering

2

0.9 s

4.3 s

3

1.9 s

9.3 s

4

3.9 s

19.3 s

5

7.8 s

39.2 s

6

15.8 s

78.9 s

7

32.0 s

2.7 min.

8

1.0 min.

5.3 min. Strong

filtering

9

2.1 min.

10.6 min.

10

4.2 min.

21.2 min.

11

8.5 min.

42.4 min.

12

17.0 min.

1.4 h

13

33.9 min.

2.8 h

Extremely strong filtering

3.2.4

Simulator Value

When the simulator value has been activated, a defined simulator value is used instead of the current measured

value.