Operation

This chapter describes handling of the device in operation.

6.1 Differential Pressure

6.1.1

Ambient Pressure

The differential pressure sensor in the device is dependent on the local absolute ambient atmospheric pressure

P

abs

. This must be communicated to the PF4 per display or HW4 software in order to attain the measuring

accuracy named in the data sheet. P

abs

can be determined with a barometer.

Note:

The "atmospheric pressure at the place of use"

QFE

is to be used and not the "normalized atmospheric

pressure"

QFF

, which is communicated in weather forecasts (see also http://de.wikipedia.org/wiki/QFE#QFE).

The ambient pressure enters the measurement as follows and the differential pressure is compensated

automatically on the basis of the ambient pressure setting:

Formula

Legend

1000

Real differential pressure [Pa]

Measured differential pressure [Pa]

Ambient pressure [hPa]

Relationship between altitude and pressure (on average):

Formula:

1

ö

.

where

1013.25