E-m-pf4-v1_00, 3 mounting, 4 differential pressure connection – ROTRONIC PF4 User Manual

Page 18

Advertising
background image

E-M-PF4-V1_00

Rotronic AG

Bassersdorf, Switzerland

Document Code

Unit

PF4 Differential Pressure Transmitter:

Instruction Manual

Instruction Manual

Document Type

Page 18 of 54

Document Title

© 2014; Rotronic AG

E-M-PF4-V1_00

4.3 Mounting

Method 1

: The PF4 comes with two screws, two wall plugs and two rubber

washers. There are two screw depressions in the bottom part of the housing

(closed on delivery). Use the template delivered with the PF4 to drill the holes

in the wall and put in the wall plugs. Place a rubber washer under each screw

head. Put a screw in each depression and push through the bottom of the

depression.

Method 2

: The mounting kit AC5002 (not included) is needed for mounting on

a DIN top-hat rail (35 mm / 1 3/8"). The mounting kit consists of two clips, which

are fastened to the back of the housing with the screws supplied.

Figure 6: Mounting kit AC5002 and drilling points for wall mounting

4.4 Differential Pressure Connection

Tubes with an internal diameter of 4 mm

can be connected. The tubes must be fastened securely so that

they do not move or vibrate during operation. This would falsify the measurement.

Configuration of differential pressure connections:

+

: Positive pressure connection

-

: Negative pressure connection

Figure 7: Differential pressure connections and E2 socket on the bottom side of the PF4

Advertising