PF4 Differential Pressure Transmitter:

Instruction Manual

Instruction Manual

Document Type

Page 45 of 54

Document Title

© 2014; Rotronic AG

E-M-PF4-V1_00

7.3.2

Adjust Pt100

Reference Value Adjustment

1. Enter a defined reference value with a Pt100 simulator or record the current nominal temperature with

a reference instrument.

2. Enter the reference value determined in MENU > Pt100 > RefValue.

3. Select

"<Adjust>"

in MENU > Pt100.

After successful adjustment, "OK" is shown in the display.

Reset Adjustment

The following procedure is used to reset adjustment to the factory setting.

1. Select

"<Reset Adjustment>"

in MENU > Pt100

After successful resetting of the adjustment, "OK" is shown in the display.