E-m-pf4-v1_00, 7maintenance, 1 service cable – ROTRONIC PF4 User Manual

Page 43: 2 service port

Advertising
background image

E-M-PF4-V1_00

Rotronic AG

Bassersdorf, Switzerland

Document Code

Unit

PF4 Differential Pressure Transmitter:

Instruction Manual

Instruction Manual

Document Type

Page 43 of 54

Document Title

© 2014; Rotronic AG

E-M-PF4-V1_00

7

Maintenance

Warning:

If only the service cable AC3009 is used to supply power to the PF4, the differential pressure shown will not be

correct because the voltage supply to the electronics is insufficient. For calibration or adjustment with the HW4

software, therefore, the power must be supplied via terminal block K2 in the bottom part of the housing.

7.1 Service Cable

Suitable service cables

 AC3006:

requires additional power supply to the device

 AC3009:

supplies the device directly with 5 VDC, differential pressure measurement will,

however, not be correct without external power supply

7.2 Service Port

The Mini-USB service port is located inside the device. The device must be opened by unscrewing the four

screws on the front of the housing in order to gain access to the Mini-USB service port.

Remark:

AC3006 / AC3009 may be used as service cable.

Figure 19: Location of the service port on the top side of the circuit board

Advertising