E-M-PF4-V1_00

Rotronic AG

Bassersdorf, Switzerland

Document Code

Unit

PF4 Differential Pressure Transmitter:

Instruction Manual

Instruction Manual

Document Type

Page 41 of 54

Document Title

© 2014; Rotronic AG

E-M-PF4-V1_00

<Adjust>

Applies change to RefValue

Temperature Adjust (1)

Display when a HC2 sensor is not connected

Temperature Adjust (2)

Display when a HC2 sensor is connected

Select

Options

RefValue

The current temperature can be changed

<Adjust>

Applies change to RefValue

Exchange Sensor

Exchange Sensor

is started after selecting Analog In in menu Probe

Information

The HC2 sensor can be substituted by an analog input (depending on order

code, for current or voltage measurement).

The action is only started after confirmation by pressing the Enter key [].

 If

the

Enter

key [] is not pressed within about 10 seconds, the device

returns to the menu Probe Information.

Exchange HC2 sensor for the analog input and press any key

If a key is not pressed within about 20 seconds, the device starts

automatically with the new settings.

The device is started with the new settings.

Summary of procedure:

MENU > Probe Information > Enter > Sensor Type HC2 > Sensor Type Analog In

> Enter > Change Sensors > Any Key