3.4.2

Simulator Value

When the simulation value has been activated, a defined simulator value is used instead of the current measured

value.

3.5 Analog Outputs

The ROTRONIC HW4 software enables free configuration and scaling of the analog outputs. The measured

values (differential pressure, humidity, temperature, calculation value) can be assigned at will to any analog

output (Out1, Out2) and the required range can also be scaled freely.

The following output ranges are available:

Type

Range

Voltage

 0…1

V

 0…5

V

 0…10

V

Current

 0…20

mA

 4…20

mA

The different variants of the analog output each have a minimal offset:

Signal Type

Maximum Offset at Start of Range

0...1 V

3 mV

0...5 V

50 mV

0...10 V

90 mV

0...20 mA

4 µA

4...20 mA

No offset