Setting a dc voltage output, Etting a, Oltage – Transmille 1000-Series User Manual

Page 36: Utput, Warning

Advertising
background image

1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 36

Setting a DC Voltage Output

Complete the following procedure to set a dc voltage output. The

and

keys

can be used to edit the entry in the event of an incorrect key press.

Warning

ENSURE THE OUTPUT DOES NOT EXCEED RATING FOR UUT INPUT

1) Ensure the calibrator’s output has been set to Standby. This can be verified

by ensuring that standby is indicated on the display, and the output standby

status LED is lit.

2) Connect the UUT to the calibrator as described for DC Voltage

measurements (see connection diagram on Page 29)

3) Select the correct range on the UUT.

4) Press

5) Press the numeric and decimal point keys to enter the required value, e.g.

56.789.

6) Press

key, depending upon the polarity of the output required (default is

positive)

7) Press the multiplier key (if required) e.g.

8) Press

9) The display will now indicate the value that has been entered below the

currently set output.