Insulation resistance calibration (option), Nsulation, Esistance – Transmille 1000-Series User Manual
Page 61: Alibration, Ption

1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 61
Insulation Resistance Calibration (Option)
The insulation resistance simulation function provides resistance calibration for
insulation testers.
1) Ensure the calibrator’s output has been set to Standby. This can be verified by
ensuring that standby is indicated on the display, and the output standby status
LED is lit.
2) Set the insulation tester to the required Insulation voltage range and connect it to
the voltage terminals of the calibrator.
3) Press the INS TEST soft key. If the INS TEST key is not visible, press the NEXT
key until the INS TEST key appears
4) Press V DOWN
or V UP to select the required range as set on the UUT
5) To change the resistance value, using the numeric keys enter the required value,
e.g. 99.8
6) Press the multiplier key (if required) i.e.
7) Press
8) The display will now indicate the value that has been entered below the
currently set output