Insulation Resistance Calibration (Option)

The insulation resistance simulation function provides resistance calibration for
insulation testers.

1) Ensure the calibrator’s output has been set to Standby. This can be verified by

ensuring that standby is indicated on the display, and the output standby status
LED is lit.

2) Set the insulation tester to the required Insulation voltage range and connect it to

the voltage terminals of the calibrator.

3) Press the INS TEST soft key. If the INS TEST key is not visible, press the NEXT

key until the INS TEST key appears

4) Press V DOWN

or V UP to select the required range as set on the UUT

5) To change the resistance value, using the numeric keys enter the required value,

e.g. 99.8


6) Press the multiplier key (if required) i.e.

7) Press

8) The display will now indicate the value that has been entered below the

currently set output