Transmille 1000-Series User Manual
Page 42

1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 42
The display will show the frequency in the bottom right hand corner of the display.
The frequency can be adjusted using the digital control or the directional keys when
the cursor is placed over the frequency.
To move the cursor from voltage setting to frequency setting, press the
key,
followed by
For protection of UUT’s, the 1000 Series calibrators high voltage outputs are
ramped, so that voltage is increased gradually. Please refer to the ‘High Voltage
Output Ramp’ section (Page 69) for further information on this function.
For safety reasons the 1000 Series calibrator is fitted with a High Voltage timeout.
Please refer to the ‘High Voltage Timeout’ section (Page 69) for further information
on this function.
For safety reasons the 1000 Series calibrator is fitted with a High Voltage current
limit. This ensures that incorrect connections, faulty UUT’s or potentially dangerous
situations are protected against. Please refer to the ‘High Voltage Current Limit’
section (Page 70) for further information on this function.