1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 49
10) Press
11) The display will now indicate the value that has been entered below the
currently set output
12) Press
to activate the calibrator output. The LED next to the
key will
light up, as well as the terminal indicator; indicating that the output is active
The output can be changed by using the
and
Using these keys will switch
to the next decade value.
Nulling the UUT
The value displayed/set on the calibrator is the value at the terminals. Therefore the
measuring instrument should be zeroed (Nulled) with the leads shorted before
connection to the calibrator.
This will remove any errors due to the resistance of the leads, especially at low
values where resistance of the test leads may be significant.