1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 49

10) Press

11) The display will now indicate the value that has been entered below the

currently set output

12) Press

to activate the calibrator output. The LED next to the

key will

light up, as well as the terminal indicator; indicating that the output is active

The output can be changed by using the

and

Using these keys will switch

to the next decade value.

Nulling the UUT

The value displayed/set on the calibrator is the value at the terminals. Therefore the
measuring instrument should be zeroed (Nulled) with the leads shorted before
connection to the calibrator.

This will remove any errors due to the resistance of the leads, especially at low
values where resistance of the test leads may be significant.