Setting frequency output – Transmille 1000-Series User Manual
Page 51

1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 51
9) Press
. The new value will replace the existing output in the middle
of the display
10) Press
to activate the calibrator output. The LED next to the
key will
light up, as well as the terminal indicator; indicating that the output is active.
The output can now be changed by using the
and
. Using these keys will
switch to the next decade value.
Capacitance is available at the voltage terminals as indicated by the terminal
indicator LED. The capacitance displayed will be the calibrated value held in the
non-volatile calibration memory for that standard. Note that this is the value
measured at 1 kHz. When measuring capacitance, Cp (parallel) should be selected
for values up to and including 1uF (where available)
It is important to ensure that the capacitance of the leads has been removed from
the measurement indicated by the UUT as this may affect