PRT Output
The calibrator can simulate PRT temperature values with R0’s (the nominal
resistance at 0ºC) of 25
100
, 250
500
and 1000
in the range -200
C to
800
C
1) Ensure the calibrator’s output has been set to Standby. This can be verified
by ensuring that standby is indicated on the display, and the output standby
status LED is lit.
2) Connect the UUT to the calibrator as described for PRT measurements (see
connection diagram on Page 32)
3) Select the correct range on the UUT
4) Press the PRT soft key. If the PRT key is not visible, press the NEXT key until
the PRT key appears
5) Press RO, the Ro Selection screen will appear
6) Select the required R0 value (typically 100) and press SELECT or
7) Press the numeric and decimal point keys to enter the required value, e.g. 30
8) Press