Dc voltage commands, Oltage, Ommands – Transmille 1000-Series User Manual
Page 74: R2/o1/s0<cr

1000 SERIES OPERATION MANUAL
TRANSMILLE LTD.
Version 1.30
Page 74
DC Voltage Commands
Function
Range
Command
DC Voltage
100mV
R1
1V
R2
10V
R3
100V
R4
1000V
R5
Standby Mode
Standby ON
S1
Standby OFF
S0
Output
Set Output
O (not zero)
The DC voltage section consists of a set of range commands which are used in
conjunction with the standby and output command. To enable a DC Voltage to be
set and an output assigned, the following command sequence should be used:
<RANGE>/<OUTPUT>/<STANDBY CONDITION><CR>
For example, to assign 1V D.C. with the output switched on, the command is:
R2/O1/S0<CR>
R2 = 1V Range (as detailed in the table above)
O1 = Set an OUTPUT of 1V
S0 = Standby OFF (i.e. output switched ON)
<CR> = Carriage Return (ASCII character 13)