1000 SERIES OPERATION MANUAL

TRANSMILLE LTD.

Version 1.30

Page 50

Setting Capacitance Output


Note: The calibrator uses standard capacitors of fixed values. The nearest available

capacitance to the entered value will be automatically selected. Complete the

following procedure to select 100nF.

Complete the following example to set an output using the simulated resistance
output.

1) Ensure the calibrator’s output has been set to Standby. This can be verified

by ensuring that standby is indicated on the display, and the output standby
status LED is lit.

2) Connect the UUT to the calibrator as described for capacitance

measurements (see connection diagram on Page 31)

3) Select the correct range on the UUT
4) Press the FREQ soft key
5) Press the numeric and decimal point keys to enter the required value, e.g.

1.234

6) Press the multiplier key (if required) e.g.

7) Press

8) The display will now indicate the value that has been entered below the