Sper Scientific 850061 pH SD Card Datalogger Kit User Manual

Page 19

Advertising
background image

- 19 -

CALIBRATION

This meter has been calibrated using
a pH probe with an ideal reading:
generates 0 mV at pH 7.00 (177.4
mV at pH 4) in an ambient
temperature of 25°C. Because your
probe may not be ideal, calibration is
necessary before first time use. For
best results, calibrate your meter
weekly. For the highest accuracy, we
recommend a two point calibration.
Begin with pH 7 (then calibrate pH 4
or 10).

To calibrate:

1. Prepare the probe for calibration

(see page 25).


2. Using a standard pH and/or ATC

probe, plug the probe plug into the
pH probe socket.

3. Press POWER to turn the meter

on.


4. For the standard pH probe only

(not ATC): Adjust the manual
temperature compensation value