Sper Scientific 850061 pH SD Card Datalogger Kit User Manual

Page 46

Advertising
background image

- 46 -

PC CONNECTION

RS232 Format: 9600, N, 8, 1

D12 &
D11

Annunciator for Display

pH =
05

mV =
18

D10

Polarity, 0 = Positive,
1 = Negative

D9

Decimal point (DP)
position from right to left
0 = No DP, 1 = 1 DP, 2 =
2 DP, 3 = 3DP

D8 to D1

Display reading, D1 =
LSD, D8 = MSD
For example: If the
display reading is 1234,
then D8 to D1 is
00001234

D0

End word

Baud rate

9600

Parity

No parity

Data bit no.

8 Data bits

Stop bit

1 Stop bit