MAINTENANCE

inaccurate reading.


If the contamination is mineral
matter, put the sensing element in
a 0.1 N HCl solution for 10
minutes. Rinse in distilled water.

If the contamination is oil or
grease, clean with a mild
detergent. Rinse in distilled water.


Upon completion of either cleaning
method, immerse the probe in a
saturated buffer solution with pH 4.01
for 15 minutes and rinse with distilled
water. After cleaning, soak the probe
in solution for at least 8 hours.

Probe Performance
The sensing element of an ORP
probe is made of a high purity metal.
Soaking the sensing element in a
solution for a long period of time may
cause slow response time and
inaccurate readings. An oxidation
reduction coating may have formed
on the surface of the sensing