MagTek InSpec 9000-2005 User Manual

Page 76

Advertising
background image

InSpec 9000-2005 Encoded Card Tester


66

Data Fault On Track (1,2,3)

Please Read Card Again.

Error detected in size data. Reread card. If problem continues, cycle tester power.

Encoded Card Tester help not available.

The InSpec 9000-2005 software was unable to find the help file in the correct location.


External RAM Failure!
or
FPGA Failure!

These messages indicate a failure in the tester electronics. Switch off the tester power.
Wait several seconds before reapplying power to the unit.



File Exists, Choose Different Date!

The user may not create two different archives using the same date. Select another date.



Incomplete Data Transmission!
or
Invalid STX byte!

or

Invalid ETX byte!
or
Invalid LRC byte

These messages indicate that the data transmitted from the tester was corrupted or
incomplete. If problem continues, cycle tester power and reread card.


Invalid File Format!

The file is not a valid saved card file. Only .CQT files can be opened by this software. If
the file was a CQT, the file may be corrupted.

Advertising