Aalborg PWE Digital User Manual
Page 13
![background image](/manuals/414121/13/background.png)
COMMUNICATIONS PARAMETERS (RS-232/RS-485):
Baud rate:
......................
9600 baud
Stop bit:
......................
1
Data bits:
......................
8
Parity: ......................
None
Flow Control:
......................
None
DISPLAY*: Optional local 2x16 characters LCD with adjustable backlight (2 lines of text).
KEY PAD*: Optional 4 push button key pad.
CE COMPLIANCE: EMC Compliance with 89/336/EEC as amended.
Emission Standard: EN 55011:1991, Group 1, Class A
Immunity Standard: EN 55082-1:1992
(* - optional feature)
Table 2.3.1 PWE Flow Meter Flow Ranges and Pressure Drop
Table 2.3.2 PWE Flow Meter wetted materials
8
FLOW RATE FOR PWE
METER
SIZES
FLOW RATE H
2
0
INLET/OUTLET
PORTS FEMALE
NPT
MAXIMUM PRESSURE DROP
[L/min]
Gal/min
Bar
PSI
PWE4
0.15-18.9 0.04-5
3/8"
1
15
PWE6
0.3-37.6
0.08-10
1/2"
1.4
20
PWE8
0.6-64.4
0.15-17
3/4"
1.4 20
PWE10
1.3-132.5
0.35-35
1"
1.4 20
WETTED MATERIALS
POLYPROPYLENE MODELS
PVDF MODELS
BODY
Polypropylene PVDF
LID
Acrylic
PVDF
PADDLE WHEEL
PVDF
Nickel Tungsten Carbide
PVDF
Zirconia Ceramic
BEARINGS
Sapphire Jewels
Sapphire Jewels
O-RINGS
EPDM
PTFE
PLATINUM RTD
316 stainless steel casing
316 stainless steel casing