Aalborg PWE Digital User Manual
Page 72
67
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Pilot
To
talizer
Sets and controls action of
the Pilot flow totalizer
.
NOTE:
Pilot T
o
talizer reading is
stored in SRAM (volatile)
memor
y. Power cycle will
reset Pilot T
o
talizer reading
to zero.
10
PT
Z (Reset to zero)
PTZ
F (start totalizer
at flow %F
.S.)
<value>
(flow %FS)
PTF:<value>
L (Limit gas
volume in
current E.U.)
<value>
(gas volume)
PTL:<value>
D (disable
totalizer)*
PT
:D
E (enable
totalizer)
PT
:E
R (read current
totalizer volume)
PTR:<value>
(in current EU)
S (setting status)
PTS: Mode, Start, Limit
Low Flow
Cut Off
Display /Change Meter Low
Flow Cut Off settings in %F
.S.
11
CO
<Cut off V
alue>
(0 to 10.0%)
CO:<Cut off V
alue>
Example: CO:2.0
No Argument
(Returns Current
Cut off V
alue
settings)
CO:<Cut off V
alue>
Example: CO:2.0
Fluid
Density
Read and set Fluid Density in
g/cm3
12
FD
(New Density
V
alue) 0.01
Density
5.0
g/cm3
FD:<value>
No Argument
(Returns Current
Density in g/cm3)
FD:<value>
Example: FD:1.000